Clear and Sanitize Procedures
Tektronix recommends that you password-protect the BIOS by setting an
Administrator password under the Security tab in the BIOS con
fi
guration.
Disable the CD-RW
Capability
Complete the following steps to disable the CD-RW capability of the instrument:
1.
Log on to the instrument as an administrator.
2.
From the Windows Start menu, go to the
Control Panel
.
3.
Select
Hardware and Sound
.
4.
Click
Device Manager
to open the Device Manger window.
5.
Expand the DVD/CD-ROM drives selection in the Device Manager window.
6.
Right-click the drive and select
Disable
.
7.
Restart the instrument to make the changes effective.
Tektronix recommends that you password-protect the Device Manger settings
by setting an Administrator password under the Security tab in the BIOS
con
fi
guration.
Disable the LAN Ethernet
Connectors
Complete the following steps to disable the LAN Ethernet connectors on the
instrument:
1.
Log on to the instrument as an administrator.
2.
From the Windows Start menu, go to the
Control Panel
.
3.
Select
Hardware and Sound
.
4.
Click
Device Manager
to open the Device Manger window.
5.
Expand the Network adapters selection in the Device Manager window.
6.
For both Gigabit Network connections, right-click and select
Disable
.
7.
Restart the instrument to make the changes effective.
Disable the RS-232 Ports
Complete the following steps to disable the RS-232 ports on the instrument:
1.
Press
Delete
on the keyboard during Boot Up to go to the BIOS con
fi
guration
menu.
2.
Go to
Advanced > NCT6776F Super IO Con
fi
guration
.
3.
Go to the
Serial Port 1 Con
fi
guration
.
4.
Set Serial Port to
Disable
.
5.
Press
Esc
to go back a menu.
6.
Go to the
Serial Port 2Con
fi
guration
.
TLA6400 Series Declassi
fi
cation and Securities Instructions
7