Adjustment Procedures
TLA5000 Series Service Manual
5- 7
Figure 5- 4: Deskew procedure probe connections
NOTE
. If any failures occur, you must complete the entire test sequence. Do not
click the Back button; otherwise the test results will be incorrect. Address the
problem and then restart the Deskew procedures.
In some cases the software will detect a failure and automatically redo the test
until it passes or times out and stops the test. You should only be concerned for a
test that continuously fails.