Performance Veri
fi
cation
10.
Check that the rising edge of the marker crosses the center horizontal graticule
line within ±2 divisions of the vertical center graticule line, as shown in the
following
fi
gure:
NOTE.
One division of displacement from graticule center corresponds to a
25 ppm time base error.
11.
Disconnect the test setup.
Check Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for all input channels.
1.
Set up the oscilloscope using the following table:
Push menu button
Select menu option
Select setting
Default Setup
—
—
Channel
1
Probe
1X
Trig Menu
Mode
Normal
Acquire
Sample
—
Source
Channel under test
Measure
Type
Pk-Pk
TDS2000C Series Oscilloscope Service Manual
4–7
Summary of Contents for TDS2001C
Page 1: ...xx TDS2000C Series Digital Storage Oscilloscopes ZZZ Service Manual P077044600 077 0446 00 ...
Page 2: ......
Page 14: ...Preface x TDS2000C Series Oscilloscope Service Manual ...
Page 15: ...Specifications ...
Page 16: ......
Page 32: ...Specifications 1 16 TDS2000C Series Oscilloscope Service Manual ...
Page 33: ...Where to Find Operating Information ...
Page 34: ......
Page 36: ...Where to Find Operating Information 2 2 TDS2000C Series Oscilloscope Service Manual ...
Page 37: ...Theory of Operation ...
Page 38: ......
Page 45: ...Performance Verification ...
Page 46: ......
Page 60: ...Performance Verification 4 14 TDS2000C Series Oscilloscope Service Manual ...
Page 61: ...Adjustment Procedures ...
Page 62: ......
Page 72: ...Adjustment Procedures 5 10 TDS2000C Series Oscilloscope Service Manual ...
Page 115: ...Replaceable Parts ...
Page 116: ......