Maintenance
Clearing the error log.
To clear the error log after
fi
xing a problem, follow these
steps:
1.
Enable the Service menus. (See page 5-2,
2.
Push the
UTILITY
button and push the
Service
option button to display
the Service menu.
3.
Push the
Service Diag.
option button to display the Diagnostics menu.
4.
Push the
Clear Error Log
option button to clear the error log of past failures.
This also sets the Power Cycle count to zero.
TDS2000C Series Oscilloscope Service Manual
6–39
Summary of Contents for TDS2001C
Page 1: ...xx TDS2000C Series Digital Storage Oscilloscopes ZZZ Service Manual P077044600 077 0446 00 ...
Page 2: ......
Page 14: ...Preface x TDS2000C Series Oscilloscope Service Manual ...
Page 15: ...Specifications ...
Page 16: ......
Page 32: ...Specifications 1 16 TDS2000C Series Oscilloscope Service Manual ...
Page 33: ...Where to Find Operating Information ...
Page 34: ......
Page 36: ...Where to Find Operating Information 2 2 TDS2000C Series Oscilloscope Service Manual ...
Page 37: ...Theory of Operation ...
Page 38: ......
Page 45: ...Performance Verification ...
Page 46: ......
Page 60: ...Performance Verification 4 14 TDS2000C Series Oscilloscope Service Manual ...
Page 61: ...Adjustment Procedures ...
Page 62: ......
Page 72: ...Adjustment Procedures 5 10 TDS2000C Series Oscilloscope Service Manual ...
Page 115: ...Replaceable Parts ...
Page 116: ......