Speci
fi
cations
Triggering System
Table 1-3: Triggering system
Characteristic
Description
Trigger Types
Edge, Video, Pulse Width
Models
Trigger source selections
TDS2001C
TDS2002C
TDS2012C
TDS2022C
Channel 1
Channel 2
External
External/5
AC Line
TDS2004C
TDS2014C
TDS2024C
Channel 1
Channel 2
Channel 3
Channel 4
External
External/5
AC Line
Trigger Source
Selection
External/5 selection attenuates the external signal by 5.
When Bandwidth Limit is selected for a channel, the bandwidth of that channel’s trigger path will also be
limited. The bandwidth of the External Trigger path is not affected by the bandwidth limit.
Horizontal Trigger
Position
The trigger position is set by the Horizontal Position knob.
500 ns minimum to 10 s maximum
Trigger Holdoff Range
The ability to set large values of Holdoff is limited by the dif
fi
culty in adjusting the Holdoff at Sec/Div
settings less than 100 ms/Div. This is because Holdoff cannot be set in Scan Mode, which begins at
100 ms/div when Trigger Mode is AUTO. By adjusting Trigger Mode to NORMAL, the Scan Mode
operation is turned off, and Holdoff can be adjusted at larger seconds/Div settings.
External Trigger Input
Impedance
1 M ±2% in parallel with 20 pF ±3 pF
300 V
RMS
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz
and above
External Trigger
Maximum Input Voltage
Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For
nonsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V should be less
than 100 ms duration and the duty factor is limited to < 44%. RMS signal level must be limited to 300 V. If
these values are exceeded, damage to the instrument may result.
Line Trigger
Characteristics
Line Trigger mode provides a source to synchronize the trigger with the AC line input.
Input Amplitude requirements: 85 V
AC
- 265 V
AC
.
Input Frequency requirements: 45 Hz - 440 Hz.
Edge Trigger
Trigger Modes
Auto, Normal
TDS2000C Series Oscilloscope Service Manual
1–7
Summary of Contents for TDS2001C
Page 1: ...xx TDS2000C Series Digital Storage Oscilloscopes ZZZ Service Manual P077044600 077 0446 00 ...
Page 2: ......
Page 14: ...Preface x TDS2000C Series Oscilloscope Service Manual ...
Page 15: ...Specifications ...
Page 16: ......
Page 32: ...Specifications 1 16 TDS2000C Series Oscilloscope Service Manual ...
Page 33: ...Where to Find Operating Information ...
Page 34: ......
Page 36: ...Where to Find Operating Information 2 2 TDS2000C Series Oscilloscope Service Manual ...
Page 37: ...Theory of Operation ...
Page 38: ......
Page 45: ...Performance Verification ...
Page 46: ......
Page 60: ...Performance Verification 4 14 TDS2000C Series Oscilloscope Service Manual ...
Page 61: ...Adjustment Procedures ...
Page 62: ......
Page 72: ...Adjustment Procedures 5 10 TDS2000C Series Oscilloscope Service Manual ...
Page 115: ...Replaceable Parts ...
Page 116: ......