Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–21
TDS 600B Test Record (cont.)
Instrument Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
TDS 600B Performance Test
Maximum
Outgoing
Incoming
Minimum
CH 3 or AX1 SIGNAL OUTPUT, 1 M
Pk-Pk
≥
80 mV
__________
__________
Pk-Pk
≤
120 mV
CH 3 or AX1 SIGNAL OUTPUT, 50
Pk-Pk
≥
40 mV
__________
__________
Pk-Pk
≤
60 mV
Probe Compensator Output Signal
Frequency (CH1 Freq).
950 Hz
__________
__________
1.050 kHz
Voltage (difference)
495 mV
__________
__________
505 mV