Performance Tests
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
1–29
Align the active cursor
over the waveform.
1
Read the
measurement
results.
2
Figure 1–4: Measurement of DC Offset Accuracy at Zero Setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING.
The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
Required
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision coaxial cables (Item 5)
Prerequisites
The oscilloscope must meet the prerequisites listed on page 1–15
See Input Channels vs. Model on page 1–2.
Check DC Voltage
Measurement Accuracy
Summary of Contents for TDS 500B
Page 4: ......
Page 6: ...Table of Contents ii TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 12: ...Preface viii TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 13: ...Performance Verification Procedures ...
Page 14: ......
Page 32: ...Performance Tests 1 18 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 102: ...Performance Tests 1 88 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 103: ...Specifications ...
Page 104: ......
Page 118: ...Nominal Traits 2 14 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 133: ......
Page 134: ......