Performance Tests
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
1–19
TDS 600B Test Record
Photocopy this and the next three pages and use them to record the performance
test results for your TDS 600B.
TDS 600B Test Record
Instrument Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
TDS 600B Performance Test
Minimum
Incoming
Outgoing
Maximum
Offset Accuracy
CH1 Offset
+1 mV
+101 mV
+1.01 V
– 1.45 mV
– 69.1 mV
– 691 mV
__________
__________
__________
__________
__________
__________
+ 1.45 mV
+ 69.1 mV
+ 691 mV
CH2 Offset
+1 mV
+101 mV
+1.01 V
– 1.45 mV
– 69.1 mV
– 691 mV
__________
__________
__________
__________
__________
__________
+ 1.45 mV
+ 69.1 mV
+ 691 mV
CH3 or AX1
+1 mV
Offset
+101 mV
+1.01 V
– 1.45 mV
– 69.1 mV
– 691 mV
__________
__________
__________
__________
__________
__________
+ 1.45 mV
+ 69.1 mV
+ 691 mV
CH4 or AX2
+1 mV
Offset
+101 mV
+1.01 V
– 1.45 mV
– 69.1 mV
– 691 mV
__________
__________
__________
__________
__________
__________
+ 1.45 mV
+ 69.1 mV
+ 691 mV
DC Voltage Measurement Accuracy (Averaged)
CH1
5 mV Vert scale setting,
–5 Div position setting, +1 V offset
+ 1.0329 V
__________
__________
+ 1.0471 V
CH1
5 mV Vert scale setting,
+5 Div position setting, –1 V offset
– 1.0471 V
__________
__________
– 1.0329V
CH1
200 mV Vert scale setting,
–5 Div position setting, +10 V offset
+ 11.4165 V
__________
__________
+ 11.7835 V
CH1
200 mV Vert scale setting,
+5 Div position setting, –10 V offset
– 11.7835 V
__________
__________
– 11.4165 V
CH1
1 V Vert scale setting,
–5 Div position setting, +10 V offset
+ 17.2425 V
__________
__________
+ 18.7575 V
CH1
1 V Vert scale setting,
+5 Div position setting, –10 V offset
– 18.7575 V
__________
__________
– 17.2425 V
CH2
5 mV Vert scale setting,
–5 Div position setting, +1 V offset
+ 1.0329 V
__________
__________
+ 1.0471 V
CH2
5 mV Vert scale setting,
+5 Div position setting, –1 V offset
– 1.0471 V
__________
__________
– 1.0329V
CH2
200 mV Vert scale setting,
–5 Div position setting, +10 V offset
+ 11.4165 V
__________
__________
+ 11.7835 V
CH2
200 mV Vert scale setting,
+5 Div position setting, –10 V offset
– 11.7835 V
__________
__________
– 11.4165 V
Summary of Contents for TDS 500B
Page 4: ......
Page 6: ...Table of Contents ii TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 12: ...Preface viii TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 13: ...Performance Verification Procedures ...
Page 14: ......
Page 32: ...Performance Tests 1 18 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 102: ...Performance Tests 1 88 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 103: ...Specifications ...
Page 104: ......
Page 118: ...Nominal Traits 2 14 TDS 500B TDS 600B TDS 700A Performance Verification and Specifications ...
Page 133: ......
Page 134: ......