background image

ECE 488: RF Circuits and Systems                                                  Spring 2005

 

CORNELL UNIVERSITY                                                                                                                © WES SWARTZ (03/8/2) 

2

particular menu was selected by the 

CH 1 MENU

 button within the Vertical Controls 

section.  The 

DC

 coupling mode means that “Direct Current” voltages are fed in (and not 

blocked as in 

AC

 

coupling).  Higher 
frequencies can be 
blocked by changing the 

BW

 (band-width) setting 

to 

ON

.  The 

10X

 probe 

setting means that the 
selected probe attenuates 
the signal by a factor of 
ten and that the readings 
on the scope will be 
adjusted to compensate 
for this factor. 
 
A second example for the 
display is shown in 
Figure 2.  Here the same 
pulse-train signal is fed 
into both channels.  Note 
the change in the time 
base. 
 
 

2. Vertical controls 

 
The vertical position and scaling of the traces for the 
two channels is controlled by the two sets of knobs 
shown in Figure 3.  The 

POSITION

 knobs adjust the 

vertical position of the zero voltage level for each 
channel.  (In extreme positions, the traces can be moved 
completely off the display.  Beware of this.)  The 

VOLTS/DIV

 (volts-per-division) knobs control the 

amount of gain applied to the signals and hence 
determine the vertical axis scales of the display.  These 
values are displayed along the bottom edge of the 
screen as mentioned above under the display section. 
 
The probe point is attached to the signal source to be 
measured, and the black wire lead is attached to a 
suitable ground (usually the ground pin of the same IC 
to minimize noise pickup).  Try adjusting these knobs 
while observing the effects on the display. 

 

 Figure 3: Vertical Controls.

Figure 2: Example of a two-trace display on the TDS 220.

Summary of Contents for TDS 220

Page 1: ...per division as indicated by CH1 2 00V in the bottom left corner of the display The digit 1 with the arrow along the left edge indicates the zero base line for the voltage of Channel 1 The time scale is set at 250 ns 250 nano seconds or 2 5X10 7 seconds per division Since there are 10 divisions from left to right the display shows a total time of 2 5 µs 2 5 micro seconds The downward pointing arro...

Page 2: ...The vertical position and scaling of the traces for the two channels is controlled by the two sets of knobs shown in Figure 3 The POSITION knobs adjust the vertical position of the zero voltage level for each channel In extreme positions the traces can be moved completely off the display Beware of this The VOLTS DIV volts per division knobs control the amount of gain applied to the signals and hen...

Page 3: ...hat the time base is properly set to avoid aliasing of high frequency signals 4 Trigger controls Triggering determines when the painting of a trace across the display begins Triggering can be automatic or forced manually depending on the settings of these controls Triggering can be repetitive for periodic waveforms or produce just a single sweep for single events Automatic triggering begins when t...

Page 4: ...top right corner of the TDS 220 will prove to be quite useful The AUTOSET button has already been mentioned under Trigger Controls but it does much more In fact whenever you first connect it to a new point on your circuit you should push AUTOSET to have the scope automatically adjust the vertical and horizontal settings as well as the trigger Then you can optimize the display using the separate ve...

Page 5: ...some superimposed noise spikes You can get a measure of this by setting the cursors via the upper knobs in the Vertical Controls section at positions that ignore the noise The automated measurement will give the extreme value from the lowest valley to the highest peak The peak to peak measurement would proceed as follows 1 Push the CURSOR button to see the Cursor menu 2 Push the top menu button to...

Page 6: ...lows 1 Push the CURSOR button to see the Cursor menu 2 Push the top menu button to select Time 3 Push the Source menu button to select CH1 4 Use the CURSOR 1 knob to place a cursor at the start of the time period to be measured as shown in Figure 8 5 Use the CURSOR 2 knob to place the remaining cursor at the end of the time period The side screen Cursor menu now shows The time at Cursor 1 The time...

Reviews: