5.
Install a probe-tip adapter on the probe (optional). This leaves your hands
free to complete the rest of the functional check. See Table 5 on page 11 for
probe-tip adapter information.
6.
Connect the probe tip to the open end of the T-BNC connector.
The spectrum analyzer displays an amplitude signal (0 dBm
±
5 dB) as
shown in Figure 8.
Figure 8: Amplitude signal
To verify the functionality of Channel B, perform the following tasks:
1.
Move the probe from Channel A to Channel B.
2.
Move the SMA cable from A OUT to B OUT.
NOTE
. The spectrum analyzer will adjust the RF amplitude offset for Channel A
only. To correct the amplitude for Channel B, see Table 6. For example, if you
are using a 5X probe on Channel B, manually apply an amplitude offset
of --14 dB.
Table 6: Probe amplitude offset for Channel B
Probe Attenuation
¶
Amplitude Offset
2.5X
--8 dB
5X
--14 dB
6.25X
--16 dB
10X
--20 dB
12.5X
--22 dB
25X
--28 dB