![Tektronix RTPA2A Instruction Manual Download Page 21](http://html1.mh-extra.com/html/tektronix/rtpa2a/rtpa2a_instruction-manual_1078175021.webp)
Getting Started
RTPA2A Instruction Manual
11
Performing a Functional Check
To verify that the TekConnect probe adapter is functioning properly, perform the
following procedure. Refer to Table 5 for a list of recommended equipment.
Table 5: Recommended equipment for functional check
Item description
Recommended example
Real-Time Spectrum Analyzer
Tektronix RSA2200A, RSA3300A, RSA3408A, or
WCA200A Series (see page 2 for spectrum analyzer
software compatibility)
USB cable
1
174-4401-xx
2
Power supply and cable
1
119-7017-xx
2
50
Ω
SMA-to-N cable
1
174-5218-xx
2
TekConnect probe
Tektronix P7000 Series
Probe tip adapter
Attaching the probe tip will leave your hands free to
perform the functional check.
Use the standard accessories recommended in the
functional check for your TekConnect probe to attach the
probe tip to a connector.
50
Ω
(M-to-M) RF coaxial cable
012-0057-xx
2
BNC (F-to-F) adapter
103-0028-xx
2
T-BNC (M-to-two F) connector
103-0030-xx
2
or equivalent
50
Ω (
BNC-type) termination
011-0049-xx
2
or equivalent
1
Included with the RTPA2A TekConnect probe adapter
.
2
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
To verify the functionality of Channel A, perform the following tasks:
1.
Set up the spectrum analyzer:
a.
Power on and wait for the boot-up process to complete, if necessary.
b.
Preset the factory defaults. For example, RSA2203A -- Press the System
key, and then press the Reset All to Factory Defaults side key.
c.
Set the center frequency to 10 MHz.
d.
Set the span to 1 MHz.
Summary of Contents for RTPA2A
Page 4: ......
Page 10: ...Environmental Considerations vi RTPA2A Instruction Manual...
Page 26: ...Getting Started 16 RTPA2A Instruction Manual...
Page 30: ...Reference 20 RTPA2A Instruction Manual...
Page 33: ...Specifications RTPA2A Instruction Manual 23 Figure 10 Insertion loss Figure 11 Return loss...