Installation
Figure 6: RF INPUT connector
Functional Check
The analyzer has a built-in calibration signal source with a frequency of 100 MHz
(RSA3408B) or 50 MHz (RSA3300B) and an amplitude of approximately
-20 dBm. Use the built-in source to perform this quick functional check to verify
that your instrument is operating correctly.
1.
Turn on the analyzer.
2.
Display the spectrum of the calibration signal:
a.
Press the
S/A
key on the front panel and then press the
Spectrum
Analyzer
side key.
b.
Press the
PRESET
key on the front panel to reset the analyzer.
c.
Press the
INPUT
key on the front panel.
d.
Press the
Signal Input Port...
side key to select
Cal100M
(RSA3408B)
or
Cal
(RSA3300B).
The spectrum of the calibration signal appears.
e.
Check that "INPUT: CAL" and "FREE RUN" are displayed in the status
indicator at the upper right of the screen. (See Figure 7.)
8
RSA3303B, RSA3308B, and RSA3408B User Manual