P4617 & P6418 Logic Analyzer Probe Instructions
11
Maintenance
The P6417 and P6418 Probes do not require scheduled or periodic maintenance.
Use the procedures listed under Functional Check to check the basic functional-
ity of the probes.
To verify that the probes meet or exceed the performance requirements for
published specifications with a compatible logic analyzer module, refer to the
TLA 700 Series Performance Verification and Adjustment Procedures and follow
the procedures listed under LA Module Performance Verification.
Functional Check
The following procedure checks the basic operation of the probes by verifying
that the probes recognize signal activity at the probe tips.
Equipment
required
Adjustment/verification fixture and power supply (refer to Replaceable
Parts for part number information)
Prerequisites
P6417 or P6418 probe connected to LA Module
Test equipment connected as shown in Figure 8
Adjustment/verification
fixture
J2
J1
J26
Fixture supply
LA module
C2/C3/CK3
Channel/Group
J3
Clock channel
J5
INT
EXT
J15
Figure 8: Probe functional verification test setup