Adjustment and Verification Procedures
P6339A Instruction Manual
27
Test Record
Photocopy the following test record and use it to record test results.
P6339A Test Record,
Page 1 of 3
Serial
Number
Procedure performed by
Date
Oscilloscope tests
Low limit
Test result
High limit
Output zero
1
10 mV
–20 mV
+20 mV
20 mV
–10 mV
+10 mV
50 mV
–5 mV
+5 mV
0.2 V
–5 mV
+5 mV
5 V
–5 mV
+5 mV
1 V
–5 mV
+5 mV
2 V
–5 mV
+5 mV
5 V
–5 mV
+5 mV
10 V
–5 mV
+5 mV
20 V
–5 mV
+5 mV
50 V
–5 mV
+5 mV
100 V
–5 mV
+5 mV
1
The high and low limits for output zero represent the acceptable
limits over time. When making the adjustments, set the
10 mV/div adjustment to 0.00 mV
±
1.0 mV and all other
adjustments to 0.00 mV
±
0.5 mV.