Adjustment and Verification Procedures
P6339A Instruction Manual
23
DC Attenuation Accuracy
1. Connect the probe as shown in Figure 12.
–
+
Digital multimeter
1103
CH 1
input
CH 1
output
50
W
precision
terminator
50
W
BNC cable
Female BNC-to-male
banana adapter
P6339A
BNC-to-probe tip adapter
013-0226-00
DC Source
Figure 12: Setup for DC attenuation accuracy
2. Set the DMM to the mV range.
3. Set all the DIP switches to the “0” position.
4. Set the DIP switches for an attenuation of 10 mV and adjust the
gain potentiometer marked 10m (top row, left) for the best
compromise that produces an output reading on the DMM within
the limits listed in Table 8 on page 24 for both positive and
negative voltage inputs.