P6133 Instruction Manual
5
Specifications
The characteristics listed in Table 1 apply to P6133 probes installed on Tektronix
2400 Series oscilloscopes unless otherwise noted.
Specifications apply when the instrument warms up for a period of at least
10 minutes in an environment that does not exceed the limits described in
Table 2 on page 6.
Table 1: Electrical characteristics
1
Characteristic
Description
Attenuation
2
10X:
±
1% at DC
Input resistance
2,3
10 M
W
±
1% at DC
Input capacitance
3
1.3 m
11.4 pF (Option 01)
10.8 pF (Option 25)
2 m
12.7 pF (Standard)
Compensation range
10 pF to 25 pF
Signal delay
Delay difference between any two probes is
less than 200 ps.
Maximum system bandwidth (–3 dB)
1.3 m
150 MHz
2 m
150 MHz
Maximum Input Voltage
4,5
300 V RMS CAT I or 300 V DC CAT I
150 V RMS CAT II or 150 V DC CAT II
100 V RMS CAT III or 100 V DC CAT III
425 V peak, @50% DF, <1 sec PW
650 V peak, @20% DF, <1 sec PW
750 V peak, @10% DF, <50
m
s PW
See Figure 7 for voltage derating curve.
1
Probe installed on Tektronix 150 MHz (or less) 2400 Series oscilloscope.
2
System characteristic.
3
Also see Figure 6.
4
As defined in EN61010-1. See Certifications and compliances in Table 4
5
RMS=Root Mean Square=rms=The square root of the average of the sum of the
squares of the instantaneous voltage in one cycle =
.
300 VDC=300V RMS, 420 V Peak (sinewave)=300 V RMS See Figure 5
.
ȍ
(fx
i
)
2
ń
n