Getting Started
P5210 Instructions
11
Functional Check
To make a simple functional check of the P5210 probe, select a
source that supplies AC line voltage and use the following
procedure. This procedure verifies a majority of the circuitry within
the probe. (For a complete performance verification, refer to the
service manual that accompanies this product.)
1. Use the installation procedure starting on page 10 to connect the
output of the P5210 probe to a measurement instrument.
2. See Table 1 on page 11. Connect the inputs, set the range, and
perform the check as each line of Table 1 indicates.
Table 1: Functional Check
Mode
Range
Setting
Input 1
(+ or –)
Input 2
(– or +)
Check
Differential
100X or
1,000X
Hot
Ground or
Neutral
Measurement instrument
displays or indicates the line
voltage
Common
Mode
100X or
1,000X
Hot (same
connection)
No signal
This completes the functional check procedure.