Performance Verification
P5205 Instruction Manual
29
Vise
Pliers
Modified BNC adapter
Plastic cover removed
Figure 12: Removing plastic covers from binding posts
Setup
WARNING.
These procedures require the application of high voltage
to the inputs of the P5205 probe. Because this adapter has exposed
metal surface, only qualified personnel should perform testing with
voltage levels that exceed 40 V rms. Only qualified personnel should
perform these procedures. All pertinent safety rules and guidelines
for elevated voltage measurements should be followed and adhered
to.
Connect the output of the probe to the vertical input of the test
oscilloscope. Let the probe warm up for 20 minutes.