Specifications
16
P5205 Instruction Manual
Table 2: Warranted Electrical Characteristics (Cont.)
Maximum Operating Input Voltage
500X differential:
±
1.3 kV (DC + peak AC)
500X common mode:
±
1 kV
RMS
CAT II
50X differential:
±
130 V (DC + peak AC)
50X common mode:
±
1 kV
RMS
CAT II
Maximum Nondestructive Input Voltage
500X and 50X differential:
±
1.5 kV (DC +
peak AC)
500X and 50X common mode:
±
1.3 kV
RMS
CAT II
Gain Accuracy
±
3% at 20
-
30
_
C, <70% RH
Temperature
Operating: 0 to 40
_
C
Nonoperating:
-
30 to +70
_
C
Humidity
Operating: <85% RH at or below +35
_
C
Nonoperating: <85% RH at or below
+60
_
C
Table 3: Certifications and Compliances
EC Declaration of
Conformity -- Low
Voltage
Compliance was demonstrated to the following specification as
listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC:
EN 61010-1/A2:1995
Safety requirements for electrical equipment for
measurement, control, and laboratory use
EN 61010-2-031:1994
Particular requirements for hand-held probe assemblies for
electrical measurement and test equipment
Approvals
UL3111-1
-
Standard for electrical measuring and test equipment
IEC 10106-2-031
-
Particular requirements for hand-held probe
assemblies for electrical measurement and test
CAN/CSA-C22.2 No. 1010.1-92 and CAN/CSA-C22.2 No.
1010.2.031-94
-
Safety requirements for electrical equipment for
measurement, control, and laboratory use
Summary of Contents for P5205
Page 1: ...Instruction Manual P5205 High Voltage Differential Probe 070 9472 01 www tektronix com ...
Page 4: ......
Page 14: ...Getting Started 2 P5205 Instruction Manual ON Figure 1 High Voltage Differential Probe ...
Page 20: ...Getting Started 8 P5205 Instruction Manual ...
Page 38: ...Maintenance 26 P5205 Instruction Manual ...
Page 40: ......
Page 54: ...Replaceable Parts 40 P5205 Instruction Manual ...