Standard Accessories
19
WARNING.
To avoid risk of electric shock or fire, do not use the P5210A test probe or
hook tip accessories on CAT III or CAT IV circuits. Refer to the derating table below.
To avoid risk of electric shock or fire, when using the P5210A test probe or hooktip
accessories with the P5200A, P5202A and P5205A probes, do not use on circuits
above 1000 V.
P5210A Probe standard accessory derating table
Combined probe and accessory common-mode voltage
and input voltage-to-earth ratings
Accessory
P5202A
1
P5200A & P5205A
1
P5210A
Extender leads
450 V CAT I
300 V CAT II
1000 V CAT II
600 V CAT III
2300 V CAT I
1000 V CAT III
Test probes
(TATP)
Small hook tips
(TASH)
Large hook tips
(TALH)
450 V CAT I
300 V CAT II
450 V CAT I
300 V CAT II
450 V CAT I
300 V CAT II
1000 V CAT II
600 V CAT II
1000 V CAT II
600 V CAT II
1000 V CAT II
600 V CAT II
2300 V CAT I
1000 V CAT II
2300 V CAT I
1000 V CAT II
2300 V CAT I
1000 V CAT II
1
The P5200A, P5202A & P5205A probes can also be used with the P5210A accessories,
but only at the reduced voltage levels listed in this table.