Installation
16
Connecting the Probe to the Circuit
WARNING.
To reduce risk of shock or fire, use only the accessories provided with
the probe. Do not exceed either the voltage rating or category ratings (for example,
CAT I, II, III, IV) of the probe or the probe accessory, whichever is the lesser of the
two. Refer to the tables below and on the following pages for the correct accessory
to use with your probe. When using multiple accessories (for example, the extender
leads & hook clips), the lowest accessory/probe ratings apply to all.
To avoid electric shock, keep your fingers behind the finger guard of the probe
accessory and away from the shaded area shown in the accessory illustrations below.
P5200A, P5202A & P5205A Probe standard accessory derating table
Combined probe and accessory common-mode voltage
and input voltage-to-earth ratings
Accessory
P5202A
P5200A & P5205A
P5210A
1
Extender leads
450 V CAT I
300 V CAT II
1000 V CAT II
600 V CAT III
2300 V CAT I
1000 V CAT III
Hook clips
(AC280-FL)
Pincer clips
(AC283-FL)
Alligator clips
(AC285-FL)
450 V CAT I
300 V CAT II
450 V CAT I
300 V CAT II
450 V CAT I
300 V CAT II
1000 V CAT II
600 V CAT III
1000 V CAT II
600 V CAT III
1000 V CAT II
600 V CAT III
1000 V CAT I
1000 V CAT III
1000 V CAT I
1000 V CAT III
1000 V CAT I
1000 V CAT III
1