Specification
P5205A
P5210A
Bandwidth
100 MHz
50 MHz
Maximum measurable
differential voltage
(DC + Peak AC)
1
50X: ±130 V
500X: ±1300 V
100X: ±560 V
1000X: ±5600 V
(V
rms
)
50X: 92 V
rms
500X: 920 V
rms
20X: 396 V
rms
200X: 2650 V
rms
2
Maximum common mode
voltage (DC + Peak AC)
3
Maximum input
voltage-to-earth (V
rms
)
3
50X: ±130 V
500X: ±1300 V
1000 V CAT II
600 V CAT III
100X: ±320 V
1000X: ±3200 V
1000 V CAT III
2300 V CAT I
CAT I Maximum Rated
Over-Voltage Transient
(OVT) (V
pk
)
4
—
2250 V
1
Beyond these limits, the output could be clipped. (See Figure 2 on page 12.)
2
This rating assumes that the common mode voltage ratings are not exceeded. (See
Figure 2 on page 12.)
3
The Common Mode and input voltage-to-earth ratings are the maximum ratings of each
input lead (+/-) to ground. (See Figure 2 on page 12.)
4
Applies to CAT I ratings only. The OVT peak is measured on top of the Peak Working
Voltage.