Performance Veri
fi
cation
Push the side-bezel
Stop
button.
Set the stop frequency to 1 MHz.
c.
Set the generator to provide a 50 kHz, 0 dBm signal.
d.
At 50 kHz, determine the test result as follows:
Note the reading on the power meter and the readout for the Reference
marker on the oscilloscope. See the following
fi
gure.
Calculate the difference between the two readings. This is the test
result.
e.
In the test record, enter the greatest result determined at this frequency
(50 kHz).
f.
Step the generator, in 100 kHz intervals, through frequencies from
100 kHz to 900 kHz. At each interval, determine the test result as follows:
Note the reading on the power meter and the readout for the Reference
marker on the oscilloscope.
Calculate the difference between the two readings. This is the test
result.
g.
In the test record, enter the greatest result determined within this frequency
range (100 kHz – 900 kHz).
h.
Change the frequency range as follows:
Change the stop frequency to 9.2 MHz.
Change the start frequency to 980 kHz.
i.
Set the generator to provide a 1 MHz, 0 dBm signal.
j.
Step the generator, in 1 MHz intervals, through frequencies from 1 MHz
to 9 MHz. At each interval, determine the test result as follows:
Note the reading on the power meter and the readout for the Reference
marker on the oscilloscope.
Calculate the difference between the two readings. This is the test
result.
k.
In the test record, enter the greatest result determined within this frequency
range (1 MHz to 9 MHz).
l.
Change the frequency range as follows:
Change the stop frequency to 92 MHz.
Change the start frequency to 9.8 MHz.
m.
Set the generator to provide a 10 MHz, 0 dBm signal.
96
MDO4000 Series Speci
fi
cations and Performance Veri
fi
cation
Summary of Contents for MDO4014-3