Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-56
2470-901-01 Rev. A /
May
2019
[:SENSe[1]]:<function>:RELative:ACQuire
This command acquires a measurement and stores it as the relative offset value.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:<function>:RELative:ACQuire
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
Details
This command triggers the instrument to make a new measurement for the selected function. This
measurement is then stored as the new relative offset level.
When you send this command, the instrument does not apply any math, limit test, or filter settings to
the measurement, even if they are set. It is a measurement that is made as if these settings are
disabled.
You must change to the function for which you want to acquire a value before sending this command.
The instrument must have relative offset enabled to use the acquired relative offset value.
After executing this command, you can use the
[:SENSe[1]]:<function>:RELative?
command
to return the last relative level value that was acquired or set.
Example
FUNC "RES"
RES:REL:ACQ
RES:REL?
RES:REL:STAT ON
Switch to resistance measurements. Acquire a relative offset
value for resistance measurements.
Query for the offset value.
Turn relative offset on.
Example output:
-5.4017E-10
Also see
[:SENSe[1]]:<function>:RELative
(on page 12-55)