2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-55
[:SENSe[1]]:<function>:RELative
This command contains the relative offset value.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
[:SENSe[1]]:<function>:RELative <n>
[:SENSe[1]]:<function>:RELative <DEF|MIN|MAX>
[:SENSe[1]]:<function>:RELative?
[:SENSe[1]]:<function>:RELative? <DEF|MIN|MAX>
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<n>
The relative offset value:
Current: -1.05 to 1.05
Resistance: –1e10 to 1e10
Voltage: -1100 to 1100
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
This command specifies the relative offset value that can be applied to new measurements. When
relative offset is enabled, all subsequent measured readings are offset by the value that is set for this
command.
You can set this value, or have the instrument acquire a value. If the instrument acquires the value,
read this setting to return the value that was measured internally.
Example
CURR:REL 0.5
CURR:REL:STAT ON
Set the relative offset for current measurements to 0.5. Enable
relative offset.
Also see
(on page 4-46)
[:SENSe[1]]:<function>:RELative:ACQuire