2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-43
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
For resistance measurements, set the averaging filter type to
moving average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
For voltage measurements, set the averaging filter type to
moving average, with a filter count of 10.
Enable the averaging filter.
Also see
[:SENSe[1]]:<function>:AVERage[:STATe]
(on page 12-43)
[:SENSe[1]]:<function>:AVERage:TCONtrol
(on page 12-44)
[:SENSe[1]]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF (0)
Usage
[:SENSe[1]]:<function>:AVERage[:STATe] <state>
[:SENSe[1]]:<function>:AVERage[:STATe]?
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<state>
The filter status; set to one of the following values:
Disable the averaging filter:
OFF
or
0
Enable the averaging filter:
ON
or
1
Details
This command enables or disables the averaging filter. When this is enabled, the reading returned by
the instrument is an averaged value, taken from multiple measurements. The settings of the filter
count and filter type for the selected measure function determines how the reading is averaged.
If you send this command without the
<function>
parameter, it sets the state of the averaging filter
for all functions.