Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-44
2470-901-01 Rev. A /
May
2019
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Also see
[:SENSe[1]]:<function>:AVERage:COUNt
(on page 12-42)
[:SENSe[1]]:<function>:AVERage:TCONtrol
(on page 12-44)
[:SENSe[1]]:<function>:AVERage:TCONtrol
This command sets the type of averaging filter that is used for the selected measure function when the
measurement filter is enabled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
REP
Usage
[:SENSe[1]]:<function>:AVERage:TCONtrol <type>
[:SENSe[1]]:<function>:AVERage:TCONtrol?
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<type>
The filter type to use when filtering is enabled; set to one of the following values:
Repeating filter:
REPeat
Moving filter:
MOVing