Section 4: Sourcing and measuring
2470 High Voltage SourceMeter Instrument Reference Manual
4-94
2470-901-01 Rev. A /
May
2019
Viewing configuration list contents
Use the following procedure to view configuration index 2 from
MyMeasList
.
Using the front panel:
1.
Press the
MENU
key.
2.
Under
Measure
, select
Config Lists
. MEASURE CONFIGURATION LISTS is displayed.
If you want to view a source configuration list, under
Source
, select
Config Lists
.
3.
Choose
Select List
. A menu of available configuration lists is displayed.
4.
Select
MyMeasList
. The configuration indexes are displayed.
5.
Select the second configuration index.
6.
Select
Index Details
.
Figure 82: Configuration list details
7.
When you are finished, select
OK
.
Deleting a configuration index
Use the following procedure to delete a configuration index from
MyMeasList
.
Using the front panel to delete a configuration index:
1.
Press the
MENU
key.
Under
Measure
, select
Config Lists
. The MEASURE CONFIGURATION LISTS screen is
displayed.