![Tektronix Keithley SourceMeter 2450 User Manual Download Page 86](http://html1.mh-extra.com/html/tektronix/keithley-sourcemeter-2450/keithley-sourcemeter-2450_user-manual_1077861086.webp)
Model 2450 Interactive SourceMeter® Instrument User's Manual
Section 9: Measuring I-V characteristics of solar cells
2450-900-01 Rev. A / June 2013
9-3
The circuit diagram for the application is shown in the following figure.
Figure 42: Model 2450 connections to a solar cell
The physical connections for the front and rear panels are shown in the following figures. Note that
you must use either the front terminals or rear terminals — you cannot mix connections.
The rear-terminal connections are triaxial. The front-panel connections are safety banana jacks.
Turn the power to the instrument off before attaching connections to the Model 2450.
The figure below shows the front-panel connections. You can make these connections with four
insulated banana cables (for example, two sets of the Keithley Instruments Model 8608 High
Performance Clip Lead Set).
Figure 43: Model 2450 4-wire connections to the front panel
The figure below shows the rear-panel connections. You can make these connections with four
triaxial cables.