Model 2450 Interactive SourceMeter® Instrument User's Manual
Section 6: Leakage current and insulation resistance
2450-900-01 Rev. A / June 2013
6-5
7. Select the button next to Source (at the bottom of the screen).
8. Enter
20 V
and select
OK
.
9. Press the
MENU
key.
10. Under Trigger, select
Templates
.
11. Select the button next to Templates and select
Duration Loop
.
12. Select the button next to Duration and enter
30s
("soak" time), and then select
OK
.
13. Select the button next to Delay and enter
0.2s
, and then select
OK
.
14. Press the
HOME
key to return to the home page.
15. Press the
TRIGGER
key to turn on the output and take measurements.
Viewing the measurements on the front-panel graph
To view the leakage current measurements on the front panel:
1. Press the
MENU
key.
2. Under Views, select
Graph
.
3. Select the
Scale
tab.
4. Under Y-Axis, select the button next to Scale Format and select
Log
.
5. Under X-Axis, select the button next to Auto Scale and select
On
.
6. Select the
Graph
tab to view the graph.
The following figure shows the front-panel graph for this example.
Figure 26: Viewing the leakage current measurement on the front-panel