Section 3: Adjustment
DAQ6510 Data Acquisition and Multimeter System Calibration and Adjustment Manual
3-10
DAQ6510-905-01 Rev. D June 2022
Rear-terminal adjustment with a 4-wire short
The following procedure provides instructions for completing a rear-terminal adjustment using a 4-
wire short. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
•
Keithley Model 7797 Calibration System
Prepare your DAQ6510 for a rear-terminal adjustment with a 4-wire
short
To prepare the DAQ6510 for rear-terminal adjustment:
1. Set the TERMINALS switch to
REAR
.
2. Install the Keithley Model 7797 Calibration System into your DAQ6510 rear slot 1.
The "Model 7797 Calibration System Instructions" can be found at
3. Allow the DAQ6510 to settle for five minutes.
Command parameters for a rear-terminal adjustment with a 4-wire
short
When calibrating your DAQ6510 for a rear-terminal adjustment with a 4-wire short, use the following
command parameters.
Send each command parameter twice. First, send the parameter using the
setup
command.
Second, send the parameter using the
execute
command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.