Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-146
707B-901-01 Rev. B / January 2015
scan.reset()
This function resets the trigger model and scan list settings to their factory default settings.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
scan.reset()
Details
When
scan.reset()
is sent, the trigger model and scan settings that are reset to the factory
defaults are:
•
scan.bypass
•
scan.mode
•
scan.scancount
•
scan.trigger.arm.stimulus
•
scan.trigger.channel.stimulus
In addition, the scan list is cleared.
Sending this function only affects the trigger model and scan list settings. To reset all instrument
settings to factory default settings, use the
reset()
command.
Example
scan.reset()
Performs a reset on the trigger model and scan
settings.
Also see
(on page 7-40)
(on page 7-135)
(on page 7-140)
(on page 7-145)
(on page 7-147)
(on page 7-153)
(on page 3-1)