Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-147
scan.scancount
This attribute sets the scan count in the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Scan reset
Create configuration script
1
Usage
scanCount
= scan.scancount
scan.scancount =
scanCount
scanCount
The present scan count value (1 to 2,000,000,000)
Details
The scan count attribute setting indicates how many times the scan list is iterated through before the
scan completes.
During a scan, the instrument iterates through the arm layer of the trigger model the specified number
of times. After performing the specified number of iterations, the instrument returns to an idle state.
Example
scan.scancount = 5
Sets the scan count to 5.
Also see
(on page 3-1)
(on page 3-1)