background image

Safety

Do not touch any object that could provide a current path to the common side of the circuit under 

test or power line (earth) ground. Always make measurements with dry hands while standing on 

a dry, insulated surface capable of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating 

instructions. If the instruments or accessories are used in a manner not speci

 ed in the 

operating instructions, the protection provided by the equipment may be impaired. 
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal 

levels are de

 ned in the speci

 cations and operating information and shown on the instrument 

panels, test 

 xture panels, and switching cards. 

When fuses are used in a product, replace with the same type and rating for continued 

protection against 

 re hazard. 

Chassis connections must only be used as shield connections for measuring circuits, NOT as 

protective earth (safety ground) connections. 
If you are using a test 

 xture, keep the lid closed while power is applied to the device under test. 

Safe operation requires the use of a lid interlock.
If a        screw is present, connect it to protective earth (safety ground) using the wire 

recommended in the user documentation.

The         symbol on an instrument means warning, risk of electric shock. Use standard safety 

precautions to avoid personal contact with these voltages.

The       symbol indicates a connection terminal to the equipment frame.

The         symbol on an instrument means caution, risk of hazard. The user must refer to the 

operating instructions located in the user documentation in all cases where the symbol is 

marked on the instrument.

The         symbol on an instrument shows that the surface may be hot. Avoid personal contact to 

prevent burns.

If this         symbol is on a product, it indicates that mercury is present in the display lamp. Please 

note that the lamp must be properly disposed of according to federal, state, and local laws.

The 

WARNING

 heading in the user documentation explains hazards that might result 

in personal injury or death. Always read the associated information very carefully before 

performing the indicated procedure.
The 

CAUTION

 heading in the user documentation explains hazards that could damage the 

instrument. Such damage may invalidate the warranty.
The 

CAUTION

 heading with the        symbol in the user documentation explains hazards that 

could result in moderate or minor injury or damage the instrument. Always read the associated 

information very carefully before performing the indicated procedure. Damage to the instrument 

may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. 
Before performing any maintenance, disconnect the line cord and all test cables. 
To maintain protection from electric shock and 

 re, replacement components in mains circuits 

— including the power transformer, test leads, and input jacks — must be purchased from 

Keithley. Standard fuses with applicable national safety approvals may be used if the rating and 

type are the same. The detachable mains power cord provided with the instrument may only be 

replaced with a similarly rated power cord. Other components that are not safety-related may 

be purchased from other suppliers as long as they are equivalent to the original component 

(note that selected parts should be purchased only through Keithley to maintain accuracy 

and functionality of the product). If you are unsure about the applicability of a replacement 

component, call a Keithley o

  ce for information. 

Unless otherwise noted in product-speci

 c literature, Keithley instruments are designed to 

operate indoors only, in the following environment: Altitude at or below 2,000 m (6,562 ft); 

temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2. 
To clean an instrument, use a cloth dampened with deionized water or mild, water-based 

cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument 

or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with 

no case or chassis (e.g., a data acquisition board for installation into a computer) should never 

require cleaning if handled according to instructions. If the board becomes contaminated and 

operation is a

 ected, the board should be returned to the factory for proper cleaning/servicing. 

Safety precaution revision as of June 2017.

Summary of Contents for KEITHLEY 4200A-SCS-PKB

Page 1: ...Model 4200A SCS PKB High Resolution I V C V Parameter Analyzer Quick Start Guide...

Page 2: ...r high transient overvoltages often associated with local AC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as category II or highe...

Page 3: ...the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that coulddamage the instrument Such damage may inva...

Page 4: ...tive humidity non condensing Storage temperature 15 C to 60 C 5 to 90 relative humidity non condensing CAUTION Carefully consider and configure the appropriate output off state source levels and compl...

Page 5: ...uick Start Guide This document It provides unpacking instructions describes basic connections reviews basic operation information and provides a quick test procedure to ensure the instrument is operat...

Page 6: ...sical damage Report any damage to the shipping agent immediately CAUTION The 4200A SCS weighs approximately 27 kg 60 lb and requires a two person lift CAUTION Do not lift the 4200A SCS using the front...

Page 7: ...e adapters four 5 CS 701 BNC tee adapters two 6 4200 MTRX ultra low noise SMU triaxial cables two 7 Safety interlock cable 8 4200 PA remote preamplifier 9 4200 TRX ultra low noise triaxial cables two...

Page 8: ...the device under test DUT to protect the operator from flying debris in the event of a system or DUT failure Make sure any test fixture connected to the system protects the operator from contact with...

Page 9: ...f the fixture is open It also enables the 4200A SCS to output higher voltages when the exterior enclosure of the fixture is closed When the safety interlock signal is asserted the switch is closed and...

Page 10: ...as the 8101 PIV included with this package are intended for applications that are less than 40 V For these applications an interlock is not needed High voltage test fixtures can be used with applicati...

Page 11: ...ixture Test connections If you are testing discrete devices you need a test fixture that is equipped with 3 lug triaxial connectors The 8101 PIV test fixture that is included with the 4200A SCS PKB al...

Page 12: ...Connect a test fixture and DUT to the 4200A SCS If you are using the supplied test fixture connect the MOSFET and test fixture as shown in this diagram...

Page 13: ...iaxial cable that connects 4200 PA to a test fixture 3 4200 MTRX 1 Ultra low noise SMU triaxial cable mini triaxial connects 4201 SMU units to a test fixture 4 TG 439 nMOSFET DUT 1 Plug into the test...

Page 14: ...ounded outlets When proper connections are made the instrument chassis is connected to power line ground through the ground wire in the power cord In addition a redundant protective earth connection i...

Page 15: ...end of the line cord into a properly grounded ac line power receptacle 3 Turn on the 4200A SCS by pushing the power switch The switch is lit when power is on The instrument starts up NOTE When first s...

Page 16: ...et to 60 Hz If the setting is wrong the 4200A SCS cannot properly reject powerline measurement noise To change powerline frequency 1 Close Clarius 2 Run KCon 3 From the System Configuration list selec...

Page 17: ......

Page 18: ...on a MOSFET The first time Clarius opens the project tree is displayed and the default project is loaded The next graphic shows the tests in this project To work with a test 1 Select vds id located un...

Page 19: ...est 2 The vds id settings should be SMU3 Provide a voltage step function for four different gate voltages 2 V 3 V 4 V and 5 V SMU2 Perform a 51 point sweep of drain voltage 0 V to 5 V at each gate vol...

Page 20: ...t only that test will run For example in the following graphic when you select only the vds id test will run In the following example one test is run To run the vds id test 1 In the project tree selec...

Page 21: ...ce of tests check the tests you want to run then check and select the device subsite or project item that contains the tests For example in the following graphic when you click all the checked tests u...

Page 22: ...can use the Analyze sheet which is similar to a Microsoft Excel spreadsheet and the graph To view vds id test results 1 Select Analyze 2 Select an icon next to View to choose the type of information...

Page 23: ...e data graph or both to an external file To save your test data 1 Select Save Data 2 On the Save Test Data As dialog box select one of the following buttons Save Sheet Select this to save only the she...

Page 24: ...Settings button To customize a graph 1 Select Graph Settings 2 To include a Legend and Title select those options from the Graph Settings menu 3 To use different colors for the graph series select Gr...

Page 25: ...I V Test...

Page 26: ...Make connections for the C V test If you are using the supplied test fixture connect the nMOSFET and test fixture as shown in this diagram...

Page 27: ...1 2 CA 447A cables 4 SMA cable male to male 100 1 5 m 3 CS 1391 tee adapters 2 SMA tee adapter female male female 4 8101 PIV Component Test Fixture 1 5 8101 309 nMOSFET with S D B connected together D...

Page 28: ...osfet This test is located under the 4terminal n fet device In the Configure pane on the next page is where you define the test In the test shown the device is connected to the CVH1 and CVL1 terminals...

Page 29: ...mosfet settings should be CVH1 Default setting which is Voltage Linear Sweep is swept from 5 V to 5 V in 0 2 V steps with a 1 MHz capacitance measurement made at each step CVL1 Default setting which i...

Page 30: ...test is run To run the cv nmosfet test 1 In the project tree select cv nmosfet to highlight it 2 On the toolbar select When the test is running Run shows two arrows circling the icon and Stop turns re...

Page 31: ...ata you can use the Analyze sheet which is similar to a Microsoft Excel spreadsheet or the graph To view cv nmosfet test results 1 Select Analyze 2 Select an icon next to View to choose the type of in...

Page 32: ...n external file To save your test data 1 Select Save Data 2 If you are saving a graph select the Graph File Format 3 On the Save Test Data As dialog box select one of the following buttons Save Sheet...

Page 33: ...graph is shown here In this example the C V curve is measured between the gate and the source drain bulk Notice that the COX CFB and VFB parameters are extracted and displayed on the screen You can cu...

Page 34: ...fixture Also check the connections from the DUT to the test fixture socket I cannot unplug the mini triaxial cable 4200A MTRX from the SMU What should I do The mini triaxial connector is a locking co...

Page 35: ...us Manual Provides comprehensive information about projects tests data analysis data calculation user libraries and customizing Clarius Hardware manuals for setup and maintenance source measure units...

Page 36: ...S and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specification and price change privileges reserved TEKTRONIX and TEK are r...

Reviews: