To connect the interlock cable:
1. Connect one end of the supplied interlock cable to
the rear panel of the 4200A-SCS. The location of the
rear-panel interlock connector is shown in the following
graphic.
2.
Connect the other end of the connector to the test fixture.
Test connections
If you are testing discrete devices, you need a test fixture
that is equipped with 3-lug triaxial connectors. The 8101-PIV
test fixture that is included with the 4200A-SCS-PKB allows
the 4200A-SCS to be connected to a discrete device.
For connections to a probe station for wafer testing, see the
4200A-SCS Learning Center .
NOTE
The outer shield of the triaxial cables connects the 8101-PIV
to the 4200A chassis, which connects to common.