Model 2600B-PM-2 Protection Module
2
071342900 / May 2015
Mechanical dimensions
Figure 2: Model 2600B-PM-2 dimensions
Electrical and physical characteristics
Specifications and characteristics are typical and subject to change without notice.
•
Maximum DC voltage (V): 220 V (Hi to Lo)
•
Maximum DC current (I): 0.5 A up to 20 V maximum (SMU panel Force Hi to Lo 2-wire mode)
•
Maximum current (I) pulse / duty cycle: 0.75 A (up to 20 V maximum)
•
Maximum on time: 30 ms
•
Minimum off time: 70 ms
•
Maximum 30% duty cycle, 10 Hz to 500 Hz cycle
•
Mounting location: Rear panel with support bracket or in line, standard Model 2601/02/11/12B connector
•
Supplied support bracket reversible for CH 1 and CH 2 use with Model 2602/12B dual-channel SMUs
•
Mean time between failures (MTBF): > 250,000 hours at continuous maximum DC or pulse power rating
The Model 2600B-PM-2 is not designed to protect against user shock risk or contact hazards and
should not be relied upon for operator shock safety protection.