Model 237-ALG-2 

PA-298 Rev. D / June 2013 

Before each use, inspect the cable, connector, and boots for wear, cracks, or breaks. Make sure the boots 
completely cover the alligator clips after making connections. 

The outer shield of the triaxial connector provides protection from voltages on the center and inner shield 
conductors. Make sure the outer shell is always connected to earth ground or a properly grounded chassis. 

Make sure both ends of the cable are properly connected before applying power. This prevents voltages from 
being exposed and connections from shorting together. 

Never touch or change connections when power is applied to the cable assembly. Always turn off all power and 
discharge all capacitors before connecting or disconnecting the cable. 

 

Mechanical characteristics 

A schematic of the cable is shown in the figure below. 

 

Figure 2: Model 237-ALG-2 schematic 

 

 

Electrical characteristics 

Working voltage:

 600 V peak center conductor to inner shield; 1,300 V peak center conductor and inner shield 

to outer shell 

Contact resistance:

 < 0.5 

 

Insulation resistance:

 10

15

 

, center conductor to inner shield (500 V test voltage, 23 °C at < 40 percent 

relative humidity) 

 

General characteristics 

Operating environment:

 32 °F to 122 °F (0 °C to 50 °C), up to 70 percent relative humidity at 

 95 °F (35 °C) 

 

 