Clear and Sanitize Procedures
1.
Power down the instrument and insert the optical module into the upper left
large slot. Note: the upper right large slot may also be used, adjusting the
channel parameter to CH3.
2.
Power up the instrument and wait for the oscilloscope application to become
active.
3.
From Table 4 for the applicable module(s), send each of the programmable
interface commands to the instrument. Note that the 80C10 contains two pairs
of wavelengths and scales.
4.
After all commands for all applicable modules have been issued to the
instrument, send the command TEKPHW:CH1:EEPROMUPDT to the
instrument. Wait at least 30 seconds before performing any other instrument
command or operation (including powering off the instrument).
Table 4: Optical Module User Wavelengths (Factory Defaults) – Sanitize Commands
Module type (all
options)
Programmable interface commands
80C01
TEKPHW:CH1:LONG “OPTWAVEL”,2,1000
TEKPHW:CH1:DOUBLE “OPTSCALE”,2,18
80C02
TEKPHW:CH1:LONG “OPTWAVEL”,2,1000
TEKPHW:CH1:DOUBLE “OPTSCALE”,2,18
80C07B
TEKPHW:CH1:LONG “OPTWAVEL”,4,1550
TEKPHW:CH1:DOUBLE “OPTSCALE”,4,780
80C08C
TEKPHW:CH1:LONG “OPTWAVEL”,4,1550
TEKPHW:CH1:DOUBLE “OPTSCALE”,4,315
80C10
TEKPHW:CH1:LONG “OPTWAVEL”,1,1550
TEKPHW:CH1:DOUBLE “OPTSCALE”,1,15
TEKPHW:CH1:LONG “OPTWAVEL”,3,1310
TEKPHW:CH1:DOUBLE “OPTSCALE”,3,10
80C11
TEKPHW:CH1:LONG “OPTWAVEL”,2,1550
TEKPHW:CH1:DOUBLE “OPTSCALE”,2,30
80C12
TEKPHW:CH1:LONG “OPTWAVEL”,3,1550
TEKPHW:CH1:DOUBLE “OPTSCALE”,3,315
DSA8200 Declassi
fi
cation and Security
5