Clear and Sanitize Procedures
Table 3: Optical Module User Wavelengths (Factory Defaults) – Query Commands
Module type (all
options)
Programmable
interface query
commands
User
wavelength
(factory
default)
User wavelength scale
(factory default)
TEKPHW:CH1:LONG?
“OPTWAVEL”,2
1000
80C01
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,2
18
TEKPHW:CH1:LONG?
“OPTWAVEL”,2
1000
80C02
TEKPHW:CH1 DOUBLE?
“OPTSCALE”,2
18
TEKPHW:CH1:LONG?
“OPTWAVEL”,4
1550
80C07B
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,4
780
TEKPHW:CH1:LONG?
“OPTWAVEL”,4
1550
80C08C
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,4
315
TEKPHW:CH1:LONG?
“OPTWAVEL”,1
1550
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,1
15
TEKPHW:CH1:LONG?
“OPTWAVEL”,3
1310
80C10
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,3
10
TEKPHW:CH1:LONG?
“OPTWAVEL”,2
1550
80C11
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,2
30
TEKPHW:CH1:LONG?
“OPTWAVEL”,3
1550
80C12
TEKPHW:CH1:DOUBLE?
“OPTSCALE”,3
315
Sanitize 80Cxx Modules to Factory Default Settings for User Wavelength(s) and
Scale(s).
The following procedure provides instructions on how to sanitize
user wavelength(s) and scale(s) to return them to their factory defaults. It
is recommended that the previous
Check 80Cxx Modules User Wavelengths
and Scales
procedure be performed
fi
rst so that the sanitize procedure is only
performed when necessary in order to reduce the risk of accidental corruption
of calibration data in the module. This procedure requires the use of one of
the programmable interfaces to the instrument (such as GPIB) and the use
of programmable interface commands. Connectivity and programmability
information is provided within the online help of the oscilloscope application.
4
DSA8200 Declassi
fi
cation and Security