Speci
fi
cations
Table 4: Trigger speci
fi
cations (cont.)
For signals having rise and fall times
≥
10 ns.
Source
Range
Any input channel
±0.20 div
Aux In (External)
±(10% of s 25 mV)
Trigger level accuracy, DC coupled,
typical
Line
Not applicable
Trigger holdoff range
20 ns minimum to 8 s maximum
Maximum serial trigger bits
128 bits
Standard serial bus interface triggering
I
2
C
Address Triggering:
7 and 10 bit user speci
fi
ed address, as well as General Call, START
byte, HS-mode, EEPROM, and CBUS
Data Trigger:
1 to 5 bytes of user speci
fi
ed data
Trigger On:
Start, Repeated Start, Stop, Missing Ack, Address, Data, or Address and Data
Maximum Data Rate:
10 Mbps
SPI
Data Trigger:
1 to 16 bytes of user-speci
fi
ed data
Trigger On:
SS Active, MOSI, MISO, or MOSI & MISO
Maximum Data Rate:
50 Mbps
CAN
Data Trigger:
1 to 8 bytes of user-speci
fi
ed data, including quali
fi
ers of equal to (=), not equal
to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Trigger On:
Start of Frame, Type of Frame, Identi
fi
er, Data, Identi
fi
er and Data, End of
Frame, Missing Ack, or Bit Stuf
fi
ng Errors
Frame Type:
Data, Remote, Error, Overload
Identi
fi
er:
Standard (11 bit) and Extended (29 bit) identi
fi
ers
Maximum Data Rate:
1 Mbps
LIN
Identi
fi
er Trigger:
6 bits of user-speci
fi
ed data, equal to (=)
Data Trigger:
1 to 8 bytes of user-speci
fi
ed data, including quali
fi
ers of equal to (=), not
equal to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or
equal to (>=), inside range, or outside range
Error:
Sync, Identi
fi
er Parity, Checksum
Trigger On:
Sync, Identi
fi
er, Data, Identi
fi
er & Data, Wakeup Frame, Sleep Frame, or Error
Maximum Data Rate:
100 kbps
16
MSO4000B and DPO4000B Series Speci
fi
cations and Performance Veri
fi