Speci
fi
cations
This chapter contains speci
fi
cations for the MSO4000B and the DPO4000B series
oscilloscopes. All speci
fi
cations are guaranteed unless noted as "typical."
Typical speci
fi
cations are provided for your convenience but are not guaranteed.
Speci
fi
cations that are marked with the
symbol are checked in
Performance
Veri
fi
cation
.
All speci
fi
cations apply to all MSO4000B and DPO4000B models unless noted
otherwise. To meet speci
fi
cations, two conditions must
fi
rst be met:
The oscilloscope must have been operating continuously for twenty minutes
within the speci
fi
ed operating temperature range. (See Table 11 on page 24.)
You must perform the Signal Path Compensation (SPC) operation described
in step 2 of the
Self Test
before evaluating speci
fi
cations. (See page 55,
Self
Test
.) If the operating temperature changes by more than 10 °C (18 °F), you
must perform the SPC operation again.
Model
Bandwidth
Analog
Channels
Sample
Rate
(1 ch)
Sample
Rate
(2 ch)
Sample
Rate
(4 ch)
Record
Length
(1 ch)
Record
Length
(2 ch)
Record
Length
(4 ch)
MSO4000B models:
MSO4104B
1 GHz
4
5 GS/s
5 GS/s
5 GS/s
20M
20M
20M
MSO4104B-L 1 GHz
4
5 GS/s
5 GS/s
2.5 GS/s
5M
5M
5M
MSO4102B
1 GHz
2
5 GS/s
5 GS/s
–
20M
20M
–
MSO4102B-L 1 GHz
2
5 GS/s
2.5 GS/s
–
5M
5M
–
DPO4000B models:
DPO4104B
1 GHz
4
5 GS/s
5 GS/s
5 GS/s
20M
20M
20M
DPO4104B-L 1 GHz
4
5 GS/s
5 GS/s
2.5 GS/s
5M
5M
5M
DPO4102B
1 GHz
2
5 GS/s
5 GS/s
–
20M
20M
–
DPO4102B-L 1 GHz
2
5 GS/s
2.5 GS/s
–
5M
5M
–
DPO4054B
500 MHz
4
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
DPO4034B
350 MHz
4
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
DPO4014B
100 MHz
4
2.5 GS/s
2.5 GS/s
2.5 GS/s
20M
20M
20M
MSO4000B and DPO4000B Series Speci
fi
cations and Performance Veri
fi
cation
1