Commands Listed in Alphabetical Order
Examples
AUTOSET EXECUTE
vertical, horizontal, and trigger controls of the oscilloscope to
automatically acquire and display the selected waveform.
AUTOSet:ENAble
Enables or disables the autoset feature. This is useful for classroom purposes
where the instructor wants the students to achieve the desired instrument settings
without the bene
fi
t of the autoset feature. This setting is not saved in setup
fi
les or
SET? or *LRN? queries. The default state is 1 (autoset enabled).
Group
Miscellaneous
Syntax
AUTOSet:ENAble {OFF|ON|0|1}
AUTOSet:ENAble?
Arguments
OFF or 0
disables autoset.
ON or 1
enables autoset.
Examples
AUTOSET:ENABLE 1
enables the front-panel
Autoset
button.
AUTOSET:ENABLE
? might return
AUTOSET:ENABLE 0
indicating that the autoset
feature is disabled.
AUXin? (Query Only)
Returns all auxiliary input parameters.
Group
Vertical
Syntax
AUXin?
AUXin:PRObe
Returns all information concerning the probe attached to auxilary input.
Group
Vertical
Syntax
AUXin:PRObe
AUXin:PRObe?
2-80
MSO4000 and DPO4000 Series Programmer Manual
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...