DMM4020
Technical Reference
Table 3-9. Adjustment Steps (cont.)
Step
Func./Command
Point
Input Signal
Description
31
VAC
1 V
1 V@1 kHz
Gain of VAC, 20 V range
32
VAC
10 V
10 V@1 kHz
33
VAC
15 V
15 V@1 kHz
34
VAC
19.99 V
19.99 V@1 kHz
35
VAC
10 V
10 V@1 kHz
Gain of VAC, 200 V range
36
VAC
100 V
100 V@1 kHz
37
VAC
150 V
150 V@1 kHz
38
VAC
199.9 V
199.9 V@1 kHz
39
VAC
37.5 V
37.5 V@1 kHz
Gain of VAC, 750 V range
40
VAC
200 V
200 V@1 kHz
41
VAC
500 V
500 V@1 kHz
42
VAC
750 V
750 V@1 kHz
Frequency – Adjust using 5520A
43
FREQ
500 Hz
1.9 V@500 Hz
Frequency
44
FREQ
1 kHz
1.9 V@1 kHz
45
FREQ
1.5 kHz
1.9 [email protected] kHz
46
FREQ
5 kHz
1.9 V@5 kHz
47
FREQ
10 kHz
1.9 V@10 kHz
48
FREQ
15 kHz
1.9 V@15 kHz
49
FREQ
50 kHz
1.9 V@50 kHz
50
FREQ
100 kHz
1.9 V@100 kHz
51
FREQ
150 kHz
1.9 V@150 kHz
52
FREQ
250 kHz
1.9 V@250 kHz
53
FREQ
500 kHz
1.9 V@500 kHz
54
FREQ
750 kHz
1.9 V@750 kHz
Ohms – Adjust using 5520A
55
OHMS
0
Ω
0
Ω
, 4 Wire
Zero point of OHMS, 200
Ω
range
56
OHMS
50
Ω
50
Ω
, 4 Wire
Gain of OHMS, 200
Ω
range
57
OHMS
100
Ω
100
Ω
, 4 Wire
58
OHMS
150
Ω
150
Ω
, 4 Wire
59
OHMS
1
Ω
1
Ω
, 4 Wire
Gain of OHMS, 2 k
Ω
range
60
OHMS
0.5 k
Ω
0.5 k
Ω
, 4 Wire
61
OHMS
1 k
Ω
1 k
Ω
, 4 Wire
62
OHMS
1.5 k
Ω
1.5 k
Ω
, 4 Wire
3-14
Summary of Contents for DMM4020
Page 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Page 6: ...DMM4020 Technical Reference iv...
Page 8: ...DMM4020 Technical Reference vi...
Page 10: ...DMM4020 Technical Reference 1 2...
Page 24: ...DMM4020 Technical Reference 1 16...
Page 26: ...DMM4020 Technical Reference 2 2...