Operating Instructions
—
DC
502
Set
the
MEASUREMENT INTERVAL switch to
MANUAL, apply
the
signal, and push the START button.
The
GATE indicator
will
light and
the progressing count
will
be displayed. Adjust the ATTEN and TRIGGER
LEVEL controls as
necessary
for a steady count. To
stop
the counting, release the
START button. The GATE
light
will
go
out and the displayed count will be held. The
displayed
count can continue when the START
button
is
depressed
again.
The counter can be
cleared to zero at any
time
by pressing
the RESET button or by moving the
MEASUREMENT
INTERVAL switch to another
position.
4-10
PRE-SCALE
INPUT.
In
the
MANUAL mode related
to
this
input,
the displayed count advances one count for
every
ten
incoming events. The incoming
events must have
transition
times and periods
suitable
for the 50 Ω pre-scale
input triggering requirements.
USING
THE COUNTER
DIRECT
INPUT Attenuation
and Trigger Level
Adjustment
Signals
to
be
counted in the DIRECT INPUT
channel
may
have a
wide variety of shapes and amplitudes, many of
which
are
unsuitable to drive the counting circuits.
Because
of
this, the signal is first passed through an attenuator, then
applied
to
a signal-shaping circuit
which converts it to
rectangular
pulses of uniform amplitude. This circuit
includes a
reference level adjustable b and — 2
volts
to which the incoming signal is compared, allowing
the
300-millivolt sensitivity window of the
signal-shaping
circuit
to
be adjusted to a convenient amplitude on the
incoming
waveform
(see Fig.
1-2). Obtaining a
steady,
reliable
reading is
dependent
upon the proper selection of
input
attenuation and proper
adjustment of the TRIGGER
LEVEL
control.
Noise Impulses
200-millivolt
Fig. 1-2.
Two examples of triggering
circuit output showing
how
proper
adjustment of TRIGGER LEVEL control can avoid
an
erroneous
count.
Generally,
the best point on a waveform for triggering
the
counter
is where the slope is steep and therefore usually
free
of
noise.
On a sine-wave signal, for example, the
steepest
slope occurs
at the zero-crossing point. Noise
pulses
or other signal components of sufficient amplitude
to
produce unwanted trigger pulses will cause an erratic or
incorrect
count. Fig. 1-2 shows the
TRIGGER LEVEL
control
adjusted to avoid
error. In critical measurement
applications,
monitor
the incoming signal with a test
oscilloscope.
Signal Connection
Coaxial cables and probes
offer very convenient means of
connecting
the signals to the front-panel input BNC
connectors.
These devices are shielded to prevent pickup of
electrostatic
interference
which
can
cause erroneous trig
gering
and a faulty
count. For the DIRECT INPUT, a X10
probe
not only reduces the
size of the signal, but also
presents
a high input
impedance to allow the circuit under
test to
perform very close to normal operating conditions.
For the
4-10 PRE-SCALE INPUT, the 50 Ω input requires
careful
impedance
matching.
If the signal must be atten
uated
to avoid exceeding the maximum input limit of 10 V,
use
50 Ω attenuator pads terminated by the 50 Ω input
impedance of the
DC 502.
Measurement
Interval
and Display Time Controls
The MEASUREMENT
INTERVAL switch
selects the
time
interval (also
called
gate time) during which the
DC 502
counts. The internal time-base circuit derives gate
times
from
an
accurate 1-MHz reference signal to make
frequency
measurements. These gate times are 0.01
s, 0.1 s,
1
s, or
10 s.
The measurement
interval selected determines
the
measurement
range and resolution. Also, the displayed
decimal
point is positioned correctly and the correct
measurement units
(MHz or
kHz) are indicated for the
corresponding switch position.
The
DISPLAY TIME control
sets the length of time a
measurement
can
be held in
the counter and displayed. The
HOLD
detent
position
allows a measurement to be held
indefinitely, or until
the counter is reset to
zero by the
front-panel RESET
button.
1-3
Summary of Contents for DC 502
Page 4: ...DC 502...
Page 24: ...39 39 Relocated on back of board SN B070000 REV MAY 1974...
Page 33: ...FIG 1 EXPLODED...