![Tektronix CSA 803C Service Manual Download Page 109](http://html1.mh-extra.com/html/tektronix/csa-803c/csa-803c_service-manual_1077966109.webp)
Performance Tests
CSA 803C Service Manual
4Ć41
This procedure verifies the correct operation and accuracy of the
CSA 803A's Prescaler trigger system.
Specifications
The specifications for Prescaler triggering are as follows:
800 mV
pĆp
at 2 GHz
600 mV
pĆp
at 3 GHz
600 mV
pĆp
at 10 GHz
Setup to Check Prescaler Triggering
CSA 803C
High Frequency Generator
Output
Power Meter
Picosecond Pulse
Labs5520A
10X PickĆoff
PRESCALER TRIGGER Input
CH 1
Power
Reference
X1
Sensor
Input
Input
10 GHz Power
Meter Sensor
SDĆ22, SDĆ24, or SDĆ26
B
10
Procedure to Check Prescaler Triggering
ăStep 1:ă
Install the SDĆ24 Sampling Head into the CH 1/CH 2 slot of the
mainframe. Set the ON/STANDBY switch to ON and let the instrument
warm up for approximately 15 minutes. Initialize the mainframe settings,
then perform the following settings and steps.
ăStep 2:ă
Connect the High Frequency Generator to the Picosecond
Pulse Labs (PSPL) 10X PickĆoff IN. Connect the PSPL
10 output to
Channel 1 of the Sampling Head.
Prescaler Triggering
Summary of Contents for CSA 803C
Page 3: ... ...
Page 4: ......
Page 8: ... ii ...
Page 16: ... x ...
Page 20: ... xiv ...
Page 22: ... xvi ...
Page 26: ... xx ...
Page 30: ... 1Ć4 ...
Page 31: ... 2Ć1 ...
Page 33: ... 2Ć3 CAUTION ...
Page 38: ... 2Ć8 ...
Page 39: ... 3Ć1 2 5 345 5 5 5 1 0 1 1 0 0 0 3 3 0 1 1 1 0 1 2 5 3 1 0 5 1 0 3 5 5 0 5 5 5 1 3 1 0 3 0 ...
Page 40: ... 3Ć2 ...
Page 74: ... 4Ć6 ...
Page 75: ... 4Ć7 0 1 0 1 0 0 1 01 0 0 0 0 0 0 1 0 1 0 0 0 0 2 2 2 2 2 2 2 2 2 2 2 2 ...
Page 80: ... 4Ć12 ...
Page 81: ... 4Ć13 ...
Page 84: ... 4Ć16 ...
Page 96: ... 4Ć28 0 2 0 2 0 0 0 Prec Voltage Source W 2 4 0 2 4 2 0 4 1 3 0 4 ...
Page 127: ... 5Ć13 0 3 3 1 1 1 1 3 1 1 3 1 0 1 2 2 2 2 2 2 1 0 2 1 3 1 1 1 0 ...
Page 137: ... 6Ć1 WARNING ...
Page 138: ... 6Ć2 CAUTION ...
Page 140: ... 6Ć4 ...
Page 141: ... 6Ć5 WARNING ...
Page 142: ... 6Ć6 CAUTION 1 0 0 0 1 0 1 0 W ...
Page 158: ... 6Ć22 CAUTION NOTE ...
Page 193: ... 6Ć57 NOTE ...
Page 196: ... 6Ć60 J1 2 3 4 ...
Page 219: ... 6Ć83 ...
Page 232: ... 6Ć96 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 ...
Page 235: ... 6Ć99 ST1 LSB ST2 ST3 ST4 MSB ST5 Test Running ST6 Test Failed Status LEDs ...
Page 247: ... 8Ć1 ...
Page 248: ... 8Ć2 ...
Page 255: ... 9Ć7 ...
Page 256: ... 9Ć8 ...
Page 257: ... ...
Page 258: ... 9Ć10 ...
Page 259: ... 9Ć11 ...
Page 260: ... 9Ć12 ...
Page 261: ... 10Ć1 1 2 1 0 0 1 1 1 1 1 1 0 1 1 1 1 2 1 1 1 0 1 1 1 2 0 0 1 1 1 1 1 1 ...
Page 266: ... 10Ć6 ...
Page 267: ... 10Ć7 ...
Page 280: ...Replaceable Parts Mechanical Parts List 10Ć20 1 2 3 4 5 6 Figure 10Ć7 ăAccessories ...