![Tektronix CSA 803C Service Manual Download Page 103](http://html1.mh-extra.com/html/tektronix/csa-803c/csa-803c_service-manual_1077966103.webp)
Performance Tests
CSA 803C Service Manual
4Ć35
ăStep 5:ă
Engage high precision calibration mode as follows:
UTILITYbutton
press
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Page to Enhanced Accuracy
touch
. . . . . . . . . . . . . . . . . . . . . . . . . .
Time Base Cal Mode
(in major menu)
touch
. . . . . . . . . . . . . . . . . .
High Prec
(in popĆup menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . . . .
Exit
(in popĆup menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sweep Rate Accuracy 2 ns/div
ăStep 6:ă
Adjust
the Sweep Rate to 2 ns/div.
Horizontal
icon
touch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Set
Main Size
to 2 ns/div
adjust upper knob
. . . . . . . . . . . . . . . . . . .
ăStep 7:ă
Adjust
the Signal Generator to the following settings:
Frequency
55 MHz
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Amplitude
1.0 V
RMS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
NOTE
Check that selected Sampling Head channel (1) indicator is flashĆ
ing. If not, select channel (1) by pressing select button.
ăStep 8:ă
Turn mainframe averaging on.
WAVEFORM button
press
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Acquire Desc
(in major menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . .
Average N
On (in popĆup menu)
touch
. . . . . . . . . . . . . . . . . . . . . . .
Select
Exit (
in the popĆup menu
)
touch
. . . . . . . . . . . . . . . . . . . . . . .
NOTE
Position the display so the sine wave is centered on the screen. To
ensure accurate period measurement, the display must be a full
cycle.
ăStep 9:ă
Obtain Period as follows:
MEASURE button
press
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement
(in major menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . .
Period
(in popĆup menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Exit Menu
(in popĆup menu)
touch
. . . . . . . . . . . . . . . . . . . . . . . . . . .
ăStep 10:ă
Examine
that the period is between 18.172 ns and 18.192 ns,
showing that the sweep rate accuracy is within .05%.
Summary of Contents for CSA 803C
Page 3: ... ...
Page 4: ......
Page 8: ... ii ...
Page 16: ... x ...
Page 20: ... xiv ...
Page 22: ... xvi ...
Page 26: ... xx ...
Page 30: ... 1Ć4 ...
Page 31: ... 2Ć1 ...
Page 33: ... 2Ć3 CAUTION ...
Page 38: ... 2Ć8 ...
Page 39: ... 3Ć1 2 5 345 5 5 5 1 0 1 1 0 0 0 3 3 0 1 1 1 0 1 2 5 3 1 0 5 1 0 3 5 5 0 5 5 5 1 3 1 0 3 0 ...
Page 40: ... 3Ć2 ...
Page 74: ... 4Ć6 ...
Page 75: ... 4Ć7 0 1 0 1 0 0 1 01 0 0 0 0 0 0 1 0 1 0 0 0 0 2 2 2 2 2 2 2 2 2 2 2 2 ...
Page 80: ... 4Ć12 ...
Page 81: ... 4Ć13 ...
Page 84: ... 4Ć16 ...
Page 96: ... 4Ć28 0 2 0 2 0 0 0 Prec Voltage Source W 2 4 0 2 4 2 0 4 1 3 0 4 ...
Page 127: ... 5Ć13 0 3 3 1 1 1 1 3 1 1 3 1 0 1 2 2 2 2 2 2 1 0 2 1 3 1 1 1 0 ...
Page 137: ... 6Ć1 WARNING ...
Page 138: ... 6Ć2 CAUTION ...
Page 140: ... 6Ć4 ...
Page 141: ... 6Ć5 WARNING ...
Page 142: ... 6Ć6 CAUTION 1 0 0 0 1 0 1 0 W ...
Page 158: ... 6Ć22 CAUTION NOTE ...
Page 193: ... 6Ć57 NOTE ...
Page 196: ... 6Ć60 J1 2 3 4 ...
Page 219: ... 6Ć83 ...
Page 232: ... 6Ć96 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 ...
Page 235: ... 6Ć99 ST1 LSB ST2 ST3 ST4 MSB ST5 Test Running ST6 Test Failed Status LEDs ...
Page 247: ... 8Ć1 ...
Page 248: ... 8Ć2 ...
Page 255: ... 9Ć7 ...
Page 256: ... 9Ć8 ...
Page 257: ... ...
Page 258: ... 9Ć10 ...
Page 259: ... 9Ć11 ...
Page 260: ... 9Ć12 ...
Page 261: ... 10Ć1 1 2 1 0 0 1 1 1 1 1 1 0 1 1 1 1 2 1 1 1 0 1 1 1 2 0 0 1 1 1 1 1 1 ...
Page 266: ... 10Ć6 ...
Page 267: ... 10Ć7 ...
Page 280: ...Replaceable Parts Mechanical Parts List 10Ć20 1 2 3 4 5 6 Figure 10Ć7 ăAccessories ...