CFG253
Performance Verification & Adjustment Procedures
Page 4 of 19
Procedure #: CP1005 Revision: A
2.
CHECK MAIN OUTPUT AMPLITUDES
a.
Set-Up:
AMPLITUDE
CW
DC OFFSET
Pushed In
RANGE
1K
FUNCTION
Triangle
VOLTS OUT
Button Out
FREQ/10
Button Out
FREQ DIAL
3.0
b.
Test Scope Set-Up:
Volts/Div
2V
Time/Div
0.2 mSec
c.
Connect MAIN OUTPUT to test scope vertical input via coax cable and 50 ohm
terminator.
d.
Check the test scope display for > 5 divisions (10V).
e.
Rotate the AMPLITUDE knob to nearly CCW.
f.
Set test scope to 10 mV/Div. Check that the display reduces to less than 5
divisions (50mV) as the AMPLITUDE knob approaches the CCW stop.
g.
Set the VOLTS OUT button in and change the test scope Volts/Div to 5mV.
h.
Check for a display of < 1 division (5mV) as the knob approaches the CCW stop.
i.
Set the test scope Volts/Div to 0.2 V. Rotate the AMPLITUDE knob fully CW.
j.
Check the display for > 5 divisions (1V).
k.
Return the VOLTS OUT button to the out position and ensure that the
AMPLITUDE is CW.