Performance Verification & Adjustment Procedures
Page 15 of 19
Procedure #: CP1005 Revision: A
5.
Square Wave Overshoot, Rise Time and Fall Time
a.
Test Scope Set-Up:
Volts/Div
5V
Time/Div
0.5 uSec
Trig Slope
+
b.
Set-Up:
AMPLITUDE
CW
DC OFFSET
Pushed In
RANGE
100K
FUNCTION
Square Wave
VOLTS OUT
Button Out
FREQ/10
Button Out
FREQ DIAL
3.0
c.
Connect MAIN OUTPUT to test scope vertical input via coax cable WITHOUT
the 50 ohm terminator.
d.
Adjust AMPLITUDE knob for a 4 division display on the test scope.
e.
Adjust C31 for minimum overshoot (will not be completely flat).
f.
Set Time/Div to 50 nSec/Div. Adjust Trig Level control as necessary to display
the rising edge of the Square Wave.
g.
Check risetime between the 10% and 90% points for < 100 nSec (2 Div).
h.
Set test scope Trig Slope to - (negative). Adjust Trig Level to view the Falling
edge of Square Wave.
i.
Check the Fall Time between the 10% and 90% points for < 100 nSec (2 Div).
6.
Set OUTPUT Levels
a.
Set-Up:
AMPLITUDE
CW
DC OFFSET
Pushed In
RANGE
1K
FUNCTION
Triangle
VOLTS OUT
Button Out
FREQ/10
Button Out
FREQ DIAL
3.0