Acquiring Piconet Packet Data
1- 12
BPA105 Bluetooth Protocol Analyzer User Manual -- COMPANY CONFIDENTIAL
1/2001
Acquisition Timeout.
In the Acquisition Setup dialog box (see Figure 1--2 on
page 1--2), use the Timeout (sec) field to set the number of seconds following
synchronization before the Bluetooth Protocol Analyzer loses synchronization if
there is no activity on the piconet. In such an event, the Bluetooth Protocol
Analyzer will display the message
Out of sync with piconet!
.
NOTE
. When you operate the Bluetooth Protocol Analyzer and Bluetooth
Neighborhood together to participant in a piconet, you must use Piconet mode.
When you operate the Bluetooth Protocol Analyzer in Independent Mode
working as a passive listener, you cannot use Bluetooth Neighborhood.
Sync Indication Bar.
When you select Independent mode as the logging mode, the
acquisition window displays a Sync Bar. See Figure 2--4.
The Sync Bar contains the following indicators:
H
Inquiry. This indicator is green when the Bluetooth Protocol Analyzer starts
the inquiry procedure during master inquiry or slave inquiry. The indicator
changes to red if the target device does not answer within a 60-second
timeout.
H
PageScan. This indicator is green when the Bluetooth Protocol Analyzer
enters the Page scan portion of the synchronization procedure. It is present
only if the slave inquiry or fake connection is selected. A timeout is not
included as part of this synchronization procedure, consequently, the user can
only stop the synchronization by clicking the toolbar Stop button.
H
Connect. This indicator is green when the Bluetooth Protocol Analyzer
enters the channel hopping sequence (the analyzer searches for first traffic on
the piconet). If no traffic is recorded, the indicator changes to red, indicating
that synchronization has failed.
H
In sync. This indicator turns green when the first packet of the channel
hopping sequence is received. If synchronization to the piconet is lost
(41-second timeout) the indicator changes to red, indicating that synchro-
nization to the piconet is lost.
Figure 1- 4: Acquisition window Sync Bar
Summary of Contents for BPA105
Page 11: ...Operating Basics ...
Page 12: ......
Page 49: ...Reference ...
Page 50: ......
Page 62: ...Reference 2 12 BPA105 Bluetooth Protocol Analyzer User Manual COMPANY CONFIDENTIAL 1 2001 ...
Page 63: ...Appendices ...
Page 64: ......
Page 79: ...Glossary ...
Page 80: ......
Page 89: ...Index ...
Page 90: ......
Page 95: ......
Page 96: ......