Calibration — Type 516
MODE
A VERTICAL
VOLTS/DIV.
VARIABLE VOLTS/DIV.
POLARITY
A ONLY
.05
CALIBRATED
NORM. DC
Set the frequency control on the signal generator for an
output frequency of 1 megahertz, and adjust the amplitude
control on the signal generator for a vertical deflection of
exactly 4 major divisions. Center the display vertically on the
crt with the A VERTICAL POSITION control.
Then increase the output frequency of the signal genera
tor to 15 megahertz. The deflection should be at least 2.8
major divisions. This corresponds to an attenuation of 3dB
or less at 15 MHz as specified (see Section 1).
6-14
Summary of Contents for 516 series
Page 5: ... Type 516 ...
Page 19: ...NOTES Type 516 ...
Page 39: ...NOTES Type 516 ...
Page 54: ...Type 516 SECTION 7 MECHANICAL PARTS LIST 7 1 ...
Page 55: ... Mechanical Parts List Type 516 FRONT 7 2 ...
Page 61: ...Mechanical Parts List Type 516 REAR 7 8 ...
Page 67: ...Mechanical Part List Type 516 LEFTSIDE 7 14 ...
Page 73: ...Mechanical Parts List Type 516 TOP 7 20 ...
Page 78: ...POWER VERTICAL AMPLIFIER CHASSIS Mechanical Parts List Type 516 7 25 ...
Page 107: ...INPUT A INPUT ft T R IG G E R IN P U T C A L O U T A TYPE S I6 OSCILLOSCOPE ...
Page 111: ... jY P O S C ILLO S C O P E F ...
Page 114: ...T 1 1 TIM IN G I69 e SWITCH ...
Page 116: ...C I R C U I T M U M tp S 2 2 5 T H R U 3 8 9 H O R IZ O N T A L A M P L IF IE R ...
Page 117: ... ...
Page 124: ...33 K UNBLANKING PULSE C IR C U IT N U M tP S 6 0 0 TH g U 8 6 9 C R T C IR C U IT TP o oA ...